SAFETY INSTRUCTIONS

  • These operating instructions describe the basic handling of the device.
  • Please refer to the operating instructions available at www.sauermanngroup.com for safe use of the product and detailed information.
  • Do not give this device to a child.
  • The magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 25 cm (10”) away from any of these devices.
  • These gas analyzers are designed to measure flue gases exhausting from combustion equipment. They are not designed for continuous operations and/or use as a control/alarm system

The Si-CA 130 and Si-CA 230 analyzers are touchscreen devices with just the on/off button:sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 2

Before Using the Flue Gas Analyzer

WARNING!

  • Be sure all connections are tight to assure accurate sampling.
  • When conducting measurements, the water trap/filter assembly MUST be in a VERTICAL position.
    Water trap OK
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 3
    Water trap NOT OK
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 4

Analyzers description

  1. Min-DIN connection for CO and CO2probe
  2. Thermocouple connection for air temperature
  3. Thermocouple connection for flue gas temperature
  4. P- connection
  5. P+ connection
  6. Gas connection
  7. USB connection
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 5

Operating temperature, protection of the instruments and information about storage

  • Operating temperature: from -5 to 45 °C (23 to 113 °F)
  • Protection: IP42

If the instrument is stored outside the operating temperature (for example in a van, a
warehouse,…) please let the analyzer acclimate to room temperature for at least 15 minutes before starting and operating the analyzer.

If the ambient temperature is lower than 10 °C (50 °F), the analyzer should be acclimated to ambient temperature before performing measurements.

First start-up

sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 6 DO NOT insert the gas probe in the flue/chimney/stack.
Keepsauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 7 pressed for 3 seconds to turn on the analyzer. sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 8

  • Tap “Start”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 9
  • Select your country then tap “Next”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 10
  • Select your language then tap “Next”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 11
  • Set the date then tap “Next”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 12
  • Set the time then tap “Next”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 13
  • Set your time zone then tap “Next”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 14
  • Select if you want your pump turned ON or OFF after Auto-Zero then tap “Save”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 15
  • Autozero is in progess.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 16
  • The probe can be inserted in the chimney.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 17
    The analyzer displays the measurements.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 18

ON / OFF

sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 19 During autozero DO NOT insert the gas probe in the chimney.sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 20

Keepsauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 21 pressed for 3 seconds to turn on the analyzer. Once the autozero is finished, insert the gas probe in the chimney and gas measurments are displayed.
sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 22

Keep sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 23pressed for 3 seconds to turn off the analyzer. Before powering off, disconnect hoses from the analyzer and empty and dry the water trap.

Stop / Restart the Pump

It is possible to turn off the pump during a gas analysis.

Tapsauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 24 . Pump is off and values are on hold.

Tap sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 25. Pump is on and measurements re-start.

Combustion Analysis, Saving & Printing

sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 26Before proceeding with the combustion analysis, remember to set the proper fuel.
sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 27

  • Tap “Settings”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 28
  • Tap “Analysis”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 29
  • Tap “Fuels”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 30
  • Tap the required fuel.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 31
  • Once the required fuel is selected, tapsauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 32 then tap “Gas analysis”.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 33
  • Measurements of the gas analysis are displayed.
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 34

You can also quickly access the fuel selection menu via the fuel shown at the top of the combustion gas analysis screen

  • To view the different measured parameters, just scroll the screen.
  • To save combustion data, first tap on the Save icon sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 35then select or create a folder and tap on Save. The Sauermann Combustion mobile phone app and/or PC software can also be used to save data.
  • To prints the results, use the Wireless Printer available as option

Software & Apps

Download the Sauermann Combustion App
sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 36  sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 37

Download the Sauermann Combustion Software
sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 38 sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 39

Download the full manual
sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 40

Monitoring Configuration Emissions Measurement on Boilers, Engines & other Applications

  • Si-CA 030
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 41
  • Si-CA 130
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 42
  • Si-CA 230
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 43
  • Wireless Printer
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 44
  • Wireless connection
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 45
  • PC and mobile Apps
    sauermann SI-CA 130 Versatile Analyzer with Touch Screen Display User Guide  & troubleshooting 46